Ripple Measurements
In the following table you will find the ripple levels that we measured on the main rails of CSP-X1200CB. According to ATX specification the limits are 120 mV (+12V) and 50 mV (5V & 3.3V).
Ripple Measurements |
---|
Test | 12 V | 5 V | 3.3 V | Pass/Fail |
20% Load | 7.2 mV | 4.8 mV | 4.8 mV | Pass |
40% Load | 9.8 mV | 5.2 mV | 5.4 mV | Pass |
50% Load | 10.8 mV | 5.6 mV | 5.4 mV | Pass |
60% Load | 12.8 mV | 6.0 mV | 6.2 mV | Pass |
80% Load | 18.8 mV | 6.8 mV | 8.0 mV | Pass |
100% Load | 24.2 mV | 9.4 mV | 11.2 mV | Pass |
Crossload 1 | 8.8 mV | 5.2 mV | 4.4 mV | Pass |
Crossload 2 | 24.2 mV | 8.4 mV | 11.6 mV | Pass |
Ripple is very low on all rails and throughout the whole load range. Even when we pulled 1200W from the unit, ripple didn't exceed 25 mV at +12V. A very low reading which shows that Andyson mastered the ripple/noise suppression, at least with this platform.
Ripple at Full Load
In the following oscilloscope screenshots you can see the AC ripple and noise that the main rails registered (+12V, 5V, 3.3V). The bigger the fluctuations on the oscilloscope's screen the bigger the ripple/noise. We set 0.01 V/Div (each vertical division/box equals to 0.01V) as standard but sometimes we are forced to use 0.02 V/Div, meaning that the fluctuations will look smaller but actually this wont be the case.
Ripple at Crossload 1
The order of images is +12V, 5V and 3.3V.
Ripple at Crossload 2
As above the order of images is +12V, 5V and 3.3V.