Ripple Measurements
In the following table you will find the ripple levels that we measured on the main rails of VP550P. According to ATX specification the limits are 120 mV (+12V) and 50 mV (5V, 3.3V and 5VSB).
Ripple Measurements Antec VP550P |
---|
Test | 12 V | 5 V | 3.3 V | 5VSB | Pass/Fail |
20% Load | 14.5 mV | 8.6 mV | 8.7 mV | 8.6 mV | Pass |
40% Load | 12.2 mV | 8.1 mV | 10.3 mV | 5.6 mV | Pass |
50% Load | 10.2 mV | 8.2 mV | 10.1 mV | 7.1 mV | Pass |
60% Load | 20.0 mV | 14.2 mV | 14.3 mV | 9.9 mV | Pass |
80% Load | 15.5 mV | 10.7 mV | 14.3 mV | 9.7 mV | Pass |
100% Load | 21.2 mV | 11.1 mV | 14.9 mV | 11.7 mV | Pass |
Crossload 1 | 11.7 mV | 19.1 mV | 16.1 mV | 9.5 mV | Pass |
Crossload 2 | 25.3 mV | 9.5 mV | 10.4 mV | 8.9 mV | Pass |
We didn't expect anything else from a Delta platform in this area. Ripple/noise suppression is outstanding, especially if we take into account that we have a budget unit for testing. Delta design and high ripple definitely is something that doesn't come along.
Ripple at Full Load
In the following oscilloscope screenshots you can see the AC ripple and noise that the main rails registered (+12V, 5V, 3.3V and 5VSB). The bigger the fluctuations on the oscilloscope's screen the bigger the ripple/noise. For all measurements we set 0.01 V/Div (each vertical division/box equals to 0.01V) as standard.
Ripple at Crossload 1
Ripple at Crossload 2